Model No.:JM-3201

Widely used in enterprise sewage, domestic sewage, surface water monitoring,

Inquiry

Detailed Description

1.Product overview:

  • New waterproof, dustproof, and moisture-proof high-quality housing (IP65) construction, with a painted surface finish.
  • The product has strong anti-interference capabilities.
  • Large-screen dot-matrix LCD display, supports Chinese and English menus (switchable), and displays multiple parameters simultaneously.
  • Capable of measuring conductivity/TDS, upper/lower limit control, and automatic range switching, enhancing the product’s measurement resolution.
  • The screen displays alarm status and simultaneously outputs ON/OFF switch signals.
  • Automatic temperature compensation function: automatic 0–100°
  • Communication functionality: Both RS485 and 4-20mA interfaces utilise optical isolation technology. The RS-485 communication interface is compatible with part of the MODBUS protocol (optional) and can be converted to RS-232. The conductivity/TDS values corresponding to the 4-20mA current output can be arbitrarily set. Supports both internal company IoT and customer-customised IoT.
  • Arbitrary hysteresis setting functionality to prevent frequent switching of the relay.
  • Customizable wide measurement range (within 0-300 mS·cm⁻¹).
  • Watchdog function: Ensures the instrument does not freeze.
  • Factory reset capability. Power failure protection > 10 years.

2.Product parameters

Measurement range 0.01 electrode 0–20.00 μS·cm-1

0.1 electrode 0–200.0 μS·cm-1

1.0 electrode 0–2000 μS·cm-1

10.0 electrode 0–20.00 mS·cm-1

Range can be customised

TDS:

0.01 electrode 0–10 ppm

0.1 electrode 0–100 ppm

1.0 electrode 0–1000 ppm

10.0 electrode 0–10,000 ppm

Range can be customised

Resolution

0.01μS·cm-1

0.1μS·cm-1

0.001 mS·cm-1

0.01 mS·cm-1

Control interface Two sets of ON/OFF relay contacts (dry contacts), divided into conductivity/TDS and temperature control signal photo-isolated outputs.
Working conditions Operating temperature range: 0°C to 60°C, relative humidity ≤90%
operating voltage 220VAC±10%、50/60Hz Size 96×96×168mm

CONTACT US TODAY!

ADD.: Road no 23-1,Pingtung city,Taiwan.

Email: sales01@jinmietw.com

Contact No.:00886-0975665338

 

 

  • About us
  • Wastewater Monitoring Instrument
  • Flue Gas Monitoring Instrument
  • News
  • Contact us
©Copyright-2025-Jinmie Enterprise Co., ltd.All rights reserved.
  • About us
  • Wastewater Monitoring Instrument
  • Flue Gas Monitoring Instrument
  • News
  • Contact us
Scroll to top Scroll to top Scroll to top